Single-Center Experience and Short-term Outcome with the JenaValve: A Second-Generation Transapical Transcatheter Aortic Valve Implantation Device

Author:

Reuthebuch Oliver1,Inderbitzin Devdas Thomas1,Ruter Florian1,Jeger Raban2,Kaiser Christoph2,Buser Peter2,Fassl Jens3,Eckstein Friedrich S.1

Affiliation:

1. Departments of Cardiac Surgery, University Hospital Basel, Basel, Switzerland.

2. Departments of Cardiology, University Hospital Basel, Basel, Switzerland.

3. Departments of Anesthesiology, University Hospital Basel, Basel, Switzerland.

Abstract

Objective We present the post-CE(Conformité Européenne)-mark single-center implantation experience and short-term outcome with the second-generation transapical JenaValve transcatheter aortic valve implantation system. Methods Patients [N = 27; 9 women; mean (SD) age, 80.3 (5.5) years] were operated on between November 2011 and August 2012. Via a transapical approach, the valve was positioned, in some cases, repositioned, and finally implanted. All data were collected during the hospital stay. Results The implantation success rate was 100%; the mean (SD) operation time was 124.7 (43.2) minutes; and the size of the implanted prosthesis was 23 mm (n = 6), 25 mm (n = 14), and 27 mm (n = 7). The in-hospital major adverse cardiac and cerebrovascular events were as follows: intraoperative resuscitation with subsequent aortic rupture (n = 1), postoperative hemorrhage needing revision (n = 1), myocardial infarction (n = 1), atrioventricular block needing a definitive pacemaker (n = 1), new-onset renal failure needing hemodialysis (n = 1), and stroke (n = 1). The 30-day mortality was 11.1% (n = 3). The mean (SD) intensive care unit/total stay was 2.2 (1.7)/11.7 (7.9) days. Postoperative echocardiography [day 6.7 (4.8)] revealed residual para-valvular leakage of trace to grade 1 in 12 patients (44.5%) and no leakage in 15 patients, with a mean (SD) transvalvular pressure gradient of 11.6 (5.6) mm Hg with significant reduction by 36.0 (17.7) mm Hg ( P = 0.0001, Wilcoxon signed rank test). Conclusions This second-generation repositionable transcatheter aortic valve implantation device could safely and successfully be implanted with a fast learning curve, significant reduction in pressure gradients, overall clinical improvement at discharge, as well as an acceptable morbidity and mortality rate in this highest-risk patient cohort.

Publisher

SAGE Publications

Subject

Cardiology and Cardiovascular Medicine,General Medicine,Surgery,Pulmonary and Respiratory Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3