Accuracy of the OMRON HEM-7361T blood pressure monitor in the differentiation between atrial fibrillation and sinus rhythm

Author:

Zhou Yi1,Xia Jia-Hui2,Zhang Wei2,Li Yan2,Wang Ji-Guang12

Affiliation:

1. Department of Epidemiology and Biostatistics, School of Public Health

2. Department of Cardiovascular Medicine, Centre for Epidemiological Studies and Clinical Trials, The Shanghai Institute of Hypertension, Ruijin Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai, China

Abstract

Objective We investigated the accuracy of the OMRON HEM-7361T automated oscillometric blood pressure (BP) monitor in the differentiation between atrial fibrillation and sinus rhythm. Methods An approximately equal number of patients with persistent atrial fibrillation and individuals with sinus rhythm were recruited from outpatients and inpatients of Ruijin Hospital, Shanghai, China. BP was measured three times consecutively with a 30-s interval with the OMRON HEM-7361T automatic electronic BP monitor for atrial fibrillation detection. A hand-held single lead electrocardiogram device was used for simultaneous electrocardiogram recordings. Results The device accurately identified atrial fibrillation in 100 (99.0%) of the 101 patients, with only 1 patient incorrectly classified as non-atrial fibrillation. The device correctly identified 99 (95.2%) of the 104 participants with sinus rhythm as non-atrial fibrillation, with five participants incorrectly classified as atrial fibrillation. The device had a positive predictive value of 95.2%, negative predictive value of 99.0%, and overall accuracy of 97.1%. Among the six misclassified participants, one with atrial fibrillation had a heart rate of 65 beats/min, and four of the five participants with sinus rhythm had cardiac arrhythmias (atrial or ventricular premature beat in one participants, sinus tachycardia in one participant, and both arrhythmias in one participant). Conclusion The OMRON HEM-7361T BP monitor is accurate in the differentiation between atrial fibrillation and sinus rhythm. Whether the device is sufficiently accurate in the differentiation between atrial fibrillation and other cardiac arrhythmias remains under investigation.

Publisher

Ovid Technologies (Wolters Kluwer Health)

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3