Validation of the JOYTECH DBP-6279B blood pressure monitor in adults and adolescents according to the AAMI/ESH/ISO universal standard(ISO 81060-2:2018 + Amd.1:2020)

Author:

Duan Baojing1,He Le2,Zhang Jianling3,Fang Lixiu4,Li Gang1

Affiliation:

1. Department of pharmacy, Hebei General Hospital

2. Cardiovascular medicine, North China University of Science and Technology, Tangshan

3. Cardiovascular medicine, Hebei Medical University Graduate School, Shijiazhuang, Hebei

4. Department of medicine, Suzhou Care-Real Medical Technology Co., Ltd., Suzhou, Jiangsu, China

Abstract

This study aimed to validate the accuracy of DBP-6279B, an automated inflationary oscillometric upper-arm blood pressure (BP) monitor, in the sitting position according to the AAMI/ESH/ISO (81060-2 : 2018 + Amd.1 : 2020) universal standard protocol. SBPs and DBPs were measured simultaneously on the same arm in 88 adults (female : male = 47 : 41) with a mean age of 56.85 years using a mercury sphygmomanometer (two observers) and a DBP-6279B device (one supervisor). The AAMI/ESH/ISO 81060-2 : 2018 and Amd.1 : 2020 universal standards for the validation of BP-measuring devices in adults and adolescents were followed. A total of 259 valid pairs of data were used in the analysis. According to Criterion 1, the mean difference of SBP between the test device (DBP-6279B) and the reference device (the mercury sphygmomanometer) was 0.75 mmHg, with a SD of 7.66 mmHg. The mean difference in DBP was 1.13 mmHg, with a SD of 6.14 mmHg. The mean difference of both SBP and DBP was less than 5 mmHg, and the SD was less than 8 mmHg, which met the requirements. According to Criterion 2, the mean difference of SBP between the test device and the reference device was 0.85 mmHg, and the SD was 6.56 mmHg, which was less than 6.88 mmHg and met the requirements. The mean difference in DBP was 1.27 mmHg, and the SD was 5.42 mmHg, which was less than 6.82 mmHg and met the requirements. DBP-6279B fulfilled the requirements of the AAMI/ESH/ISO universal standard (ISO 81060-2 : 2018 + Amd.1 : 2020); hence, it can be recommended for both clinical and self/home BP measurement in adults and adolescents.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Advanced and Specialized Nursing,Assessment and Diagnosis,Cardiology and Cardiovascular Medicine,General Medicine,Internal Medicine