Biomechanical Analysis of Multilevel Posterior Cervical Spinal Fusion Constructs

Author:

Murphy Timothy P.12,Colantonio Donald F.12,Le Anthony H.3,Fredericks Donald R.12,Schlaff Cody D.12,Holm Erik B.4,Sebastian Arjun S.5,Pisano Alfred J.12,Helgeson Melvin D.12,Wagner Scott C.12

Affiliation:

1. Department of Orthopaedics

2. Department of Surgery

3. DoD-VA Extremity Trauma and Amputation Center of Excellence, Walter Reed National Military Medical Center, Bethesda, MD

4. Uniformed Services University of the Health Sciences, Bethesda, MD

5. Department of Orthopedic Surgery, Mayo Clinic, Rochester, MN

Abstract

Study Design: Controlled Laboratory Study Objective: To compare multilevel posterior cervical fusion (PCF) constructs stopping at C7, T1, and T2 under cyclic load to determine the range of motion (ROM) between the lowest instrumented level and lowest instrumented-adjacent level (LIV-1). Summary of Background Data: PCF is a mainstay of treatment for various cervical spine conditions. The transition between the flexible cervical spine and rigid thoracic spine can lead to construct failure at the cervicothoracic junction. There is little evidence to determine the most appropriate level at which to stop a multilevel PCF. Methods: Fifteen human cadaveric cervicothoracic spines were randomly assigned to 1 of 3 treatment groups: PCF stopping at C7, T1, or T2. Specimens were tested in their native state, following a simulated PCF, and after cyclic loading. Specimens were loaded in flexion-extension), lateral bending, and axial rotation. Three-dimensional kinematics were recorded to evaluate ROM. Results: The C7 group had greater flexion-extension motion than the T1 and T2 groups following instrumentation (10.17±0.83 degree vs. 2.77±1.66 degree and 1.06±0.55 degree, P <0.001), and after cyclic loading (10.42±2.30 degree vs. 2.47±0.64 degree and 1.99±1.23 degree, P<0.001). There was no significant difference between the T1 and T2 groups. The C7 group had greater lateral bending ROM than both thoracic groups after instrumentation (8.81±3.44 degree vs. 3.51±2.52 degree, P=0.013 and 1.99±1.99 degree, P=0.003) and after cyclic loading. The C7 group had greater axial rotation motion than the thoracic groups (4.46±2.27 degree vs. 1.26±0.69 degree, P=0.010; and 0.73±0.74 degree, P=0.003) following cyclic loading. Conclusion: Motion at the cervicothoracic junction is significantly greater when a multilevel PCF stops at C7 rather than T1 or T2. This is likely attributable to the transition from a flexible cervical spine to a rigid thoracic spine. Although this does not account for in vivo fusion, surgeons should consider extending multilevel PCF constructs to T1 when feasible. Level of Evidence: Not applicable.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Neurology (clinical),Orthopedics and Sports Medicine,Surgery

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3