Abstract
Purpose:
Although the sensor-to-head distance is theoretically known to affect the signal strength in magnetoencephalography (MEG), these values have not been reported for a whole-head MEG system in a large population. We measured the distance and signal strength in 996 patients with epilepsy.
Methods:
The MEG sensor array consisted of 102 measurement sites, each of which had two gradiometers and one magnetometer. The sensor–head distance was defined as the minimum distance between each site and a set of digitized scalp points. For the signal strength, we calculated the root-mean-square of the signal values in each sensor over a recording of 4 minutes. For analyses at the individual and sensor levels, these values were averaged over the sensors and patients, respectively. We evaluated the correlation between distance and signal strength at both individual and sensor levels. At the sensor level, we investigated regional differences in these measures.
Results:
The individual-level analysis showed only a weak negative correlation between the sensor–head distance and the signal strength. The sensor-level analysis demonstrated a considerably negative correlation for both gradiometers and magnetometers. The sensor–head distances showed no significant differences between the regions, whereas the signal strength was higher in the temporal and occipital sensors than in the frontal and parietal sensors.
Conclusions:
Sensor–head distance was not a definitive factor for determining the magnitude of MEG signals in individuals. Yet, the distance is important for the signal strength at a sensor level. Regional differences in signal strength may need to be considered in the analysis and interpretation of MEG.
Funder
National Institutes of Health
Publisher
Ovid Technologies (Wolters Kluwer Health)