Performance evaluation of the sterilization process with Bowie & Dick test and biological indicator in the quality control of a blood bank in Peru

Author:

Lagos-Palomino Lesdy1,Rueda-Torres Lenin2,Sanchez-Holguin Gloria1,Soncco-Llulluy Fernando3,Rosales-Rimache Jaime45ORCID

Affiliation:

1. Universidad Alas Peruanas, Lima, Perú

2. Centro Nacional de Salud Ocupacional y Protección del Ambiente para la Salud, Instituto Nacional de Salud, Lima, Perú

3. Carrera de Medicina Humana, Universidad Científica del Sur, Lima, Perú

4. Vicerectorado de Investigación, Universidad Privada Norbert Wiener, Lima, Perú

5. .Escuela Profesional de Tecnología Médica, Universidad Continental, Lima, Perú.

Abstract

Managing contaminated waste in blood banks is a crucial process that must end with the safe disposal of blood products. In this sense, practical methods and indicators must be used to guarantee the operability of autoclave equipment and the sterility of treated waste to mitigate the risk of biological exposure and hospital biosafety. We designed a cross-sectional study to evaluate the performance the autoclaving process with Bowie & Dick test and biological indicator at the Hemotherapy and Blood Bank Unit of the Cayetano Heredia Hospital in Peru. Fifty autoclaving processes were carried out independently for the Bowie & Dick tests and biological indicators based on Geobacillus stearothermophilus spores. Autoclaving was programmed at 134°C for 3.5 minutes for the Bowie & Dick test, while, for the biological indicator, it was programmed at 121°C for 5 minutes. Both in the presence of contaminated waste. The autoclaving process evaluated by the Bowie & Dick test had a compliance rate of 80% (CI95: 66.3%–90.0%), while, by the biological indicator, 90% (CI95: 78.2%–96.7%). We did not find significant differences between the results of both tests (P = .689; Fisher exact test). The compliance rate in the autoclaving process within the blood bank of a Peruvian Hospital is acceptable; however, there are actions for continuous improvement, especially at the vacuum level in the autoclaving process.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

General Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3