Attempted retrieval of guidewire fragment using the twisting wire technique causes coronary perfusion: Case report

Author:

Zhang Zi-Wei1,Pan Xiao-Juan1,Zou Ji-Hong1,Qi Feng1ORCID

Affiliation:

1. Department of center for coronary heart disease, Fu Wai Yunan Cardiovascular Hospital, Kunming, China.

Abstract

Rationale: Guidewire fracture is one of the biggest risks of percutaneous coronary intervention, twisting wire technique is very useful for retrieving the fractured wire, but the potential risks have been inadequately reported. Herein, we present a case of retrieval of guidewire fragments using the twisting wire technique that causes coronary perfusion. Patient concerns: A 37-year-old male patient was admitted to our hospital for elective percutaneous coronary intervention of the left circumflex coronary artery. Clinical findings: For chronic total occlusion of the distal left circumflex coronary artery, antegrade recanalization by wire escalation, and parallel wire techniques were attempted. However, we shockingly found that the BMW guidewire, anchored in the right coronary artery, spontaneously fractured from the proximal right coronary artery, and a lengthy fragment of the guidewire remained in the coronary. Diagnoses, interventions, and outcomes: Many attempts were made to retrieve the remnant guidewire including the twisting wire technique, which leads to the perforation of the coronary. Outcomes: Finally, percutaneous retrieving procedures were stopped in favor of surgical extraction via a small coronary arteriotomy. This procedure was successful. Lessons: To the best of our knowledge, the present case is the first reported spontaneous fracture of the guidewire. Leaving such a lengthy remnant guidewire in the artery, or leaving stenting over the wire, would impose a high risk of coronary thrombosis, perforation, and embolization. Yet, the perforation of the artery that occurred in this case, which could have had life-threatening consequences, resulted from our attempts to retrieve the guidewire using the twisting wire technique.

Publisher

Ovid Technologies (Wolters Kluwer Health)

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3