Utility of Second-look Ultrasonography in Distinguishing BI-RADS 4 Calcifications Detected on Mammography: An observational study

Author:

Cheng Sheng1ORCID,Zhu Lin1ORCID,Lin Ning1,He Muzhen2,Yu Yingbin2,Lin Mengbo3,Zhang Hui3

Affiliation:

1. Department of Ultrasonography, Shengli Clinical Medical College of Fujian Medical University, Fujian Provincial Hospital, Fuzhou University Affiliated Provincial Hospital, Fuzhou, Fujian Province, China

2. Department of Radiology, Shengli Clinical Medical College of Fujian Medical University, Fujian Provincial Hospital, Fuzhou University Affiliated Provincial Hospital, Fuzhou, Fujian Province, China

3. Department of Surgical Oncology, Shengli Clinical Medical College of Fujian Medical University, Fujian Provincial Hospital, Fuzhou University Affiliated Provincial Hospital, Fuzhou, Fujian Province, China.

Abstract

This study aimed to assess the utility of second-look ultrasonography (US) in differentiating breast imaging reporting and data system (BI-RADS) 4 calcifications initially detected on mammography (MG). BI-RADS 4 calcifications have a wide range of positive predictive values. We hypothesized that second-look US would help distinguish BI-RADS 4 calcifications without clinical manifestations and other abnormalities on MG. This study included 1622 pure BI-RADS 4 calcifications in 1510 women (112 patients with bilateral calcifications). The cases were randomly divided into training (85%) and testing (15%) datasets. Two nomograms were developed to differentiate BI-RADS 4 calcifications in the training dataset: the MG-US nomogram, based on multifactorial logistic regression and incorporated clinical information, MG, and second-look US characteristics, and the MG nomogram, based on clinical information and mammographic characteristics. Calibration of the MG-US nomogram was performed using calibration curves. The discriminative ability and clinical utility of both nomograms were compared using the area under the receiver operating characteristic curve (AUC) and the decision analysis curve (DCA) in the test dataset. The clinical information and imaging characteristics were comparable between the training and test datasets. The bias-corrected calibration curves of the MG-US nomogram closely approximate the ideal line for both datasets. In the test dataset, the MG-US nomogram exhibited a higher AUC than the MG nomogram (0.899 vs 0.852, P = .01). DCA demonstrated the superiority of the MG-US nomogram over the MG nomogram. Second-look US features, including ultrasonic calcifications, lesions, and moderate or marked color flow, were valuable for distinguishing BI-RADS 4 calcifications without clinical manifestations and other abnormalities on MG.

Publisher

Ovid Technologies (Wolters Kluwer Health)

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3