Online Circular Contrast Perimetry: A Comparison to Standard Automated Perimetry

Author:

Meyerov Joshua1ORCID,Deng Yuanchen1,Busija Lazar23,Bigirimana Deus24,Skalicky Simon E.23

Affiliation:

1. St Vincent’s Hospital Clinical School, University of Melbourne, Melbourne, Vic., Australia

2. Glaucoma Investigation and Research Unit, the Royal Victorian Eye and Ear Hospital, Melbourne, Vic., Australia

3. Department of Surgery Ophthalmology, University of Melbourne, Melbourne, Vic., Australia

4. ChM Clinical Ophthalmology, University of Edinburgh, United Kingdom

Abstract

Purpose: The aim was to validate and compare the diagnostic accuracy of a novel 24-degree, 52-loci online circular contrast perimetry (OCCP) application to standard automated perimetry (SAP). Design: Prospective cohort study. Methods: Two hundred and twenty participants (125 normal controls, 95 open angle glaucoma patients) were included. Agreement, correlation, sensitivity, specificity, and area under receiver operating curves (AUC) were compared for parameters of OCCP, SAP, and optical coherence tomography (OCT) for the retinal nerve fiber layer and macular ganglion cell complex inner plexiform layer. Results: Pointwise sensitivity for OCCP was greater than SAP by 1.02 log units (95% CI: 0.95–1.08); 95% limits of agreement 0.860 to 1.17. Correlation and agreement for global indices and regional zones between OCCP and SAP were strong. OCCP mean deviation (MD) AUC was 0.885±0.08, similar to other instruments’ parameters with the highest AUC: SAP MD (0.851±0.08), OCT retinal nerve fiber layer inferior thickness (0.908±0.07), OCT ganglion cell complex inner plexiform layer inferior thickness (0.849±0.08), P>0.05. At best cutoff, OCCP MD sensitivity/specificity were comparable to SAP MD (90/74 vs 94/65%). Conclusions: OCCP demonstrates similar perimetric sensitivities to SAP and similar AUC to SAP and OCT in distinguishing glaucoma patients from controls. OCCP holds promise as a glaucoma surveillance and screening tool, with the potential to be utilized for in-clinic and at-home perimetry and expand community testing.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Ophthalmology,General Medicine

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3