Abstract

Background: The mechanisms of fracture in magnetically controlled growing rods (MCGRs) and the risk factors associated with this are poorly understood. This retrospective analysis of explanted MCGRs aimed to add understanding to this subject. Methods: From our cohort of over 120 retrieved MCGRs, we identified 7 rods that had fractured; all were single-rod constructs, retrieved from 6 patients. These were examined and compared with 15 intact single-rod constructs. Retrieval and fractographic analyses were used to determine the failure mode at the fracture site and the implant's functionality. Cobb angle, degree of rod contouring, and the distance between anchoring points were computed on anteroposterior and lateral radiographs. Results: 5/7 versus 3/15 rods had been inserted after the removal of a previously inserted rod, in the fractured versus control groups. All fractured rods failed due to bending fatigue. Fractured rods had greater rod contouring angles in the frontal plane (P = 0.0407) and lateral plane (P = 0.0306), and greater distances between anchoring points in both anteroposterior and lateral planes (P = 0.0061 and P = 0.0074, respectively). Conclusions: We found all failed due to a fatigue fracture and were virtually all single rod configurations. Fracture initiation points corresponded with mechanical indentation marks induced by the intraoperative rod contouring tool. Fractured rods had undergone greater rod contouring and had greater distances between anchoring points, suggesting that it is preferable to implant double rod constructs in patients with sufficient spinal maturity to avoid this complication. Clinical Relevance: Level III.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Orthopedics and Sports Medicine,General Medicine,Pediatrics, Perinatology and Child Health

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3