Affiliation:
1. Department of Radiology, Stanford University, Stanford, CA 94305, USA.
2. Department of Biomedical Engineering, University of Southern California, Los Angeles, CA 90089, USA.
Abstract
Volumetric ultrasound imaging has the potential for operator-independent acquisition and enhanced field of view. Panoramic acquisition has many applications across ultrasound; spanning musculoskeletal, liver, breast, and pediatric imaging; and image-guided therapy. Challenges in high-resolution human imaging, such as subtle motion and the presence of bone or gas, have limited such acquisition. These issues can be addressed with a large transducer aperture and fast acquisition and processing. Programmable, ultrafast ultrasound scanners with a high channel count provide an unprecedented opportunity to optimize volumetric acquisition. In this work, we implement nonlinear processing and develop distributed beamformation to achieve fast acquisition over a 47-centimeter aperture. As a result, we achieve a 50-micrometer −6-decibel point spread function at 5 megahertz and resolve in-plane targets. A large volume scan of a human limb is completed in a few seconds, and in a 2-millimeter dorsal vein, the image intensity difference between the vessel center and surrounding tissue was ~50 decibels, facilitating three-dimensional reconstruction of the vasculature.
Publisher
American Association for the Advancement of Science (AAAS)
Cited by
3 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献