Agreement between cardiac output estimation with a wireless, wearable pulse decomposition analysis device and continuous thermodilution in post cardiac surgery intensive care unit patients

Author:

Khanna Ashish K.1,Garcia Julio O2,Saha Amit K1,Harris Lynnette1,Baruch Martin3,Martin R. Shayn1

Affiliation:

1. Atrium Health Wake Forest Baptist Medical Center

2. Wake Forest School of Medicine

3. Caretaker Medical

Abstract

Abstract Purpose Pulse Decomposition Analysis (PDA) uses integration of the systolic area of a distally transmitted aortic pulse as well as an arterial stiffness estimate to compute cardiac output. We sought to assess agreement of cardiac output (CO) estimation between continuous pulmonary artery catheter (PAC) guided thermodilution (CO-CCO) and a wireless, wearable non-invasive technology that utilizes the Pulse Decomposition Analysis (CO-PDA) method in post-operative cardiac surgery patients. Methods Post processed CO-CCO measurements were compared with CO-PDA measurements in prospectively enrolled adult cardiac surgical intensive care unit patients. Both calibrated and uncalibrated CO-PDA values were compared for accuracy with CO-CCO via a Bland-Altman analysis considering repeated measurements and a concordance analysis with a 10% exclusion zone. Results 259.7 hours of monitoring data from 41 patients matching 15583 data points was analyzed. Mean CO-CCO was 5.55 L/min, while mean values for the uncalibrated CO-PDA were 5.73 L/min (bias +- precision 0.79 ± 1.11 L/min; limits of agreement − 1.43 to 3.01 L/min) and for the calibrated CO-PDA were 5.60 L/min (bias +- precision 0.22 ± 1.03 L/min; limits of agreement − 1.86 to 2.26 L/min). The percentage errors were 37.5% (uncalibrated) and 35.0% (calibrated). CO-CCO correlation with CO-PDA was moderate (0.54 and 0.67, respectively) and concordance was 0.83. Conclusion Cardiac output measured with the CO-PDA agrees with the commonly used CO-CCO PAC continuous thermodilution method. These devices are not interchangeable when using a 30% error threshold though the CO-PDA offers a non-invasive, wireless solution to complement and extend hemodynamic monitoring within and outside the ICU.

Publisher

Research Square Platform LLC

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3