Paediatric National Diagnostic Reference Levels of Computed Tomography Examination in Ghana

Author:

Shirazu Issahaku1,Sackey Theophilus A.1,Edziah Raymond2,Mensah Y. B3,Tiburu Elvis K.4,Schandorf C.4

Affiliation:

1. Ghana Atomic Energy Commission

2. University of Cape Coast

3. Korle-Bu Teaching Hospital

4. University of Ghana

Abstract

Abstract Introduction: The aim of the study was to establish and propose the utilisation of National Diagnostic Reference Levels (DRLs) for DLP (dose length product) and CTDIvol (volume weighted computed tomography dose index) for routine computed tomography (CT) for paediatric examinations. Additionally, the study estimated signal-to-noise ratio and lifetime attributable radiation risk. The materials used include multi-detector computed tomography machine, quality control equipment, Head and Body phantoms and MeVisLab workstation. Methods: The methodology involved estimation of CTDIVOL, DLP, using a minimum of 20 patients per CT facility, imaging parameters for the head, chest, and abdomen-pelvis CT examinations. In all 300 images of randomly selected paediatric patients undergoing CT scans of head, chest, and abdomen-Pelvis from five centres. 200 images which met the selection criterion were used for the analysis. Results: The measured median and upper quartile CTDIvol for head were 6.86 and 7.33 mGy, chest was 6.98 and 6.70 mGy, abdomen-pelvis were 4.71 and 5.28 mGy respectively. While DLP for head was 1103.00 and 1249.58 mGy-cm, chest was 978.86 and 1250.42 mGy-cm, abdomen-pelvis were 565.85 and 787.05 mGy-cm. Conclusion: The results of the CTDI and DLP values were comparable to international standard values. It also recommends the use of the established CTDI and DLP as reference values for future intercomparison of data from similar studies.

Publisher

Research Square Platform LLC

Reference16 articles.

1. Radiological protection and safety in medicine. ICRP Publication 73;International Commission on Radiological Protection;Ann ICRP,1996

2. Diagnostic reference levels in medical imaging: review and additional advice;International Commission on Radiological Protection;ICRP Supporting Guidance 2 Ann ICRP,2001

3. Patient Dose Assessment and Optimisation of Pelvic Radiography with Computed Radiography Systems;Abbeyquaye D;Radiat Prot Dosimetry

4. Paediatric head trauma: changes in use of computed tomography in emergency departments in the United States over time;Blackwell CD;Ann Emerg Med,2007

5. Paediatric CT imaging trends in Australia;Brady Z;J Med Imaging Radiat Oncol,2011