Comparison of Intraocular Pressure measurements obtained by Goldmann Applanation Tonometer, Corvis ST and a conventional non-contact airpuff tonometer in eyes with previous myopic refractive surgery and correlation with corneal biomechanical parameters

Author:

Karmiris Efthymios1,Tsiogka Anastasia2,Stavrakas Panagiotis3,Tsiripidis Konstantinos1,Papakonstantinou Evangelia4,Chalkiadaki Evangelia1

Affiliation:

1. Hellenic Airforce General Hospital

2. General Military Hospital of Athens

3. University of Patras

4. National and Kapodistrian University of Athens

Abstract

Abstract Purpose To assess the agreement among four types of intraocular pressure (IOP) measurements: IOP obtained by Goldmann applanation tonometer (IOP-GAT),IOP obtained by an air-puff tonometer (Nidek NT-510) (IOP-NCT), the non-corrected IOP obtained by the Corneal Visualization Scheimpflug Technology (IOP-Corvis) and the biomechanically corrected IOP obtained by the Corvis ST (bIOP-Corvis) in eyes having undergone myopic refractive surgery and correlate their differences with corneal biomechanical parameters. Methods This prospective, observational study recruited 104 post-myopic refractive surgery eyes. Each patient underwent IOP evaluation via GAT, Nidek NT-510 and Corvis ST. Difference in IOP readings was assessed by ANOVA analysis. Tonometer intermethod agreement was assessed by the Bland-Altman method. The difference between the IOP measurements was correlated against corneal thickness(CCT), age, gender, type of refractive surgery and corneal biomechanics with mixed effects linear regression analysis. Results bIOP-Corvis showed the highest values (14.77 ± 2.3mmHg), followed by IOP-NCT (13.95 ± 2.6mmHg), IOP-GAT (13.46 ± 2.9mmHg) and IOP-Corvis (12.56 ± 3.1mmHg). There were statistically significant differences in IOP measurements among all the ANOVA pairwise comparisons. Bland Altman analysis revealed a notable bias (all p < 0.01) among bIOP-Corvis and IOP-NCT, bIOP-Corvis and IOP-Corvis, bIOP-Corvis and IOP-GAT, IOP-NCT and IOP-Corvis, IOP-NCT and IOP-GAT and IOP-Corvis and IOP-GAT. We observed a strong correlation of the difference between bIOP-Corvis and IOP-NCT with patient age (p < 0.001), CCT (p < 0.001), of the difference between bIOP-Corvis and IOP-Corvis, with type of refractive surgery (p = 0.012), age(p = 0.050), CCT(p < 0.001), the stiffness parameter at first applanation(SP-A1) (p = 0.008), Ambrozio Relational Thickness horizontal (ARTh) (p < 0.001), of the difference between bIOP-Corvis and IOP-GAT, with CCT (p < 0.001), ARTh (p < 0.001), Deformation Amplitude Ratio (DA Ratio) (p = 0.035), of the difference between IOP-NCT and IOP-Corvis, with type of refractive surgery (p = 0.002), CCT (p = 0.031), SP-A1 (p = 0.014), ARTh (p < 0.001), of the difference between IOP-NCT and IOP-GAT with ARTh (p < 0.001) and of the difference between IOP-Corvis and IOP-GAT with SP-A1 (p = 0.027). Conclusion The different IOP values obtained by Corvis ST, NCT and GAT tonometers do not have a steady relationship in their measurements between them and as a result cannot be considered interchangeable in post-myopic refractive surgery eyes. These differences from each couple of IOP measurements are correlated with corneal biomechanics,CCT and age. Our results can arguably speculate that the smaller the degree of myopic correction the lower the difference between the tonometer’s readings.

Publisher

Research Square Platform LLC

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3