Resolution Enhancement for Compressed Sensing at 1.5T MRI through Deep Learning: Application to Quantitative Volume Measurement of Human Brain Structures

Author:

Suwannasak Atita1,Angkurawaranon Salita1,Sangpin Prapatsorn2,Chatnuntawech Itthi3,Wantanajittikul Kittichai1,Yarach Uten1

Affiliation:

1. Chiang Mai University

2. Philips (Thailand) Ltd

3. National Nanotechnology Center (NANOTEC)

Abstract

Abstract Background Compressed sensing (CS) is available for clinical 1.5T scanners with limited reduction factor due to a small channel of receiver coil. Typically, low-resolution (LR) with CS is able to further reduce scan time. However, LR images may be insufficient for clinical diagnosis. Recently, Deep Learning (DL) approaches have demonstrated the ability to generate high-resolution (HR) MR images from LR images. Thus, this study investigated the possibility of using LR-CS and DL-based super-resolution technique for brain volume measurement (BVM) application at 1.5T MRI. Materials and Methods For model training, human brain volumes acquired with 3D-TFE-T1W were incorporated, in which LR images were generated from the original HR images with 2x sub-sampled strategy. After data augmentation, pairs of LR and HR images were used for training 3D Residual Dense Net (RDN). For model testing, LR CS-3D-TFE-T1W images were acquired using 1.5T MRI with one minute scan time. Normalized Root-Mean-Square Error (NRMSE), Peak Signal to Noise Ratio (PSNR), and Structural Similarity (SSIM) were used for model evaluation. BVMs were performed using Freesurfer’s software. Wilcoxon signed rank test, Pearson’s correlation, and effect size were used for statistical analysis. Results The results showed that DL-SR model is able to synthesize HR images from LR images, in which no significant differences between DL-SR and actual HR (p < 0.01) reported by NRMSE (0.051 vs 0.059), PSNR (25.885 vs 24.679), SSIM (0.961 vs 0.951). For volumetric assessments, there were no significant differences between DL-SR and actual HR images (p > 0.01, Pearson’s correlation > 0.90) at seven subcortical regions. Conclusions The combination of LR CS-MRI and DL-SR can effectively address the issue of prolonged scan time in 3D MRI scans while preserving the image quality and the accuracy of brain volume measurements.

Publisher

Research Square Platform LLC

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3