Affiliation:
1. National and Kapodistrian University of Athens School of Health Sciences: Ethniko kai Kapodistriako Panepistemio Athenon
2. National and Kapodistrian University of Athens
Abstract
Abstract
Purpose
The goal of this preclinical study is to assess the functionality, technical feasibility, and safety of a new vascular robotic LIBERTYR 3 System, in the catheterization of vascular targets using a range of guidewires and microcatheters.
Material and Methods
An anesthetized pig served as an arterial model for the robotic device (LIBERTYR3; Microbot Medical Ltd, Yoqneam, IL). The primary efficacy endpoint was the evaluation of its capability to catheterize predetermined distal arterial branches in the liver, kidneys, and mesenteric arteries (technical success), under fluoroscopy guidance. The primary safety endpoint was the occurrence of angiographic acute catheterization-related complications (dissection, thrombosis, embolism, perforation). The catheterizations were conducted by two interventional radiologists that present different work experience in endovascular procedures (18 and 2 years respectively), using a variety of microcatheters and wires. Various procedural parameters such as functionality, practicality, ease of use, and time required for catheterization, were evaluated, and recorded.
Results
All pre-determined arteries were successfully catheterized (100% technical success), by both operators. No angiographic acute complications occurred. The catheters and wires were manipulated using the remote portable console in an effortless manner that maintained a high level of accuracy. Mean time for selective catheterization was 126 ± 51 seconds. The robot's conversion function to manual operation was successfully demonstrated.
Conclusion
Robotic navigation and catheterization of selected target arteries were accomplished without observable vascular damage, suggesting that the LIBERTYR 3 robotic system is a reliable and safe tool for robotic-assisted endovascular navigation. Further experimental studies are required to evaluate safety and efficacy prior to introduction into clinical practice.
Publisher
Research Square Platform LLC
Reference18 articles.
1. Minimally invasive gastrointestinal surgery: From past to the future;Rudiman R;Annals of Medicine and Surgery,2021
2. A short history of robotic surgery;Lane T;The Annals of The Royal College of Surgeons of England,2018
3. Occupational Health Risks in Cardiac Catheterization Laboratory Workers;Andreassi MG,2016
4. Review of emerging surgical robotic technology;Peters BS;Surg Endosc,2018
5. Proper Shielding Technique in Protecting Operators and Staff From Radiation Exposure in the Fluoroscopy Environment;Klein LW;J Invasive Cardiol,2021