Affiliation:
1. Max-Planck-Institute for Chemistry
Abstract
Abstract
Allan-Werle-plots are an established tool in infrared absorption spectroscopy to quantify temporal stability, maximum integration time and best achievable precision of a measurement instrument. In field measurements aboard a moving platform, however, long integration times reduce time resolution and smooth atmospheric variability. A high accuracy and time resolution are necessary as well as an appropriate estimate of the measurement uncertainty. In this study, Allan-Werle-plots of calibration gas measurements are studied to analyze the temporal characteristics of a Quantum Cascade Laser Absorption Spectrometer (QCLAS) instrument for airborne operation. Via least-squares fitting the individual noise contributions can be quantified and different dominant regimes can be identified. Through simulation of data according to the characteristics from the Allan-Werle-plot, the effects of selected intervals between in-flight calibrations can be analyzed. An interval of 30 min is found sufficient for successful drift correction during ground operation. The linear interpolation of the sensitivity increases the accuracy and lowers the measurement uncertainty from 1.1 % to 0.2 %. Airborne operation yields similar results during segments of stable flight but additional flicker and sinusoidal contributions. Simulations verify an appropriate interval of 30 min in airborne operation. The expected airborne measurement uncertainty is in the range 0.8 % to 2.4 %.
Publisher
Research Square Platform LLC
Cited by
2 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献