Affiliation:
1. Iwate Medical University
2. Nakadori General Hospital
3. Iwate Prefectural Ofunato Hospital
Abstract
Abstract
Purpose
The region behind the coronary calcium could not be visualized by intravascular ultrasound (IVUS) because of acoustic shadow. However, some pathological studies have shown that IVUS delineated the vessel border behind thin coronary calcium sheets. This study aimed to reveal whether recent IVUS can visualize the region behind thin coronary calcium sheets.
Methods
Using 534 cross-sectional optical frequency domain images (OFDI), including severe calcified coronary lesions, calcium sheet thickness was measured by every 1°. Accordingly, the visibility of the vessel border behind the coronary calcium sheet was evaluated using 60-MHz IVUS images, which were longitudinally linked with OFDI ones. After carefully coordinating with the axial position, the association between the IVUS-derived permeability of the coronary calcium sheet and calcium thickness was evaluated.
Results
The maximum and mean calcium thickness by OFDI was 0.88 ± 0.39 and 0.62 ± 0.30 mm, respectively. By 60-MHz IVUS, 12.1% of the coronary calcium sheets had permeable segments. Comparing between OFDI and IVUS images, 48.6% of the coronary calcium sheets with maximum thickness ≤ 0.3 mm were sometimes permeable by 60-MHz IVUS, whereas most > 0.5-mm thick calcium sheets were impermeable. In the receiver operating characteristic curve analysis, the best cutoff values for the maximum and mean thickness of permeable calcium were 0.48 and 0.31 mm, respectively.
Conclusion
Thus, 60-MHz IVUS can occasionally visualize the region behind a thin coronary calcium sheet. When using 60-MHz IVUS, this finding may be a predictive marker of calcium sheet with a thickness of < 0.5 mm.
Publisher
Research Square Platform LLC