In-silico evaluation of the effect of set-up errors on dose delivery during mouse irradiations with a cell irradiator-based collimator system

Author:

Entezam Amir1ORCID,Fielding Andrew2,Ratnayake Gishan3,Fontanarosa Davide4

Affiliation:

1. QUT IHBI: Queensland University of Technology Institute of Health and Biomedical Innovation

2. Queensland University of Technology Centre for Medical and Health Physics: Queensland University of Technology Institute of Health and Biomedical Innovation

3. University of Queensland - Saint Lucia Campus: The University of Queensland

4. Queensland University of Technology IHBI: Queensland University of Technology Institute of Health and Biomedical Innovation

Abstract

Abstract Set-up errors are a problem for pre-clinical irradiators that lack imaging capabilities. The aim of this study was to investigate the impact of the potential set-up errors on the dose distribution for a mouse with a xenographic tumour irradiated with a Gammacell irradiator equipped with an in-house lead collimator with 10 mm diameter apertures. The EGSnrc Monte-Carlo (MC) code was used to simulate the potential errors caused by displacements of the mouse in the irradiation setup. The impact of the simulated set-up displacements on the dose delivered to the xenographic tumour and surrounding organs was assessed. MC dose calculations were performed in a Computed Tomography (CT) derived model of the mouse for the reference position of the tumour in the irradiation setup. The errors were added into the CT data and then the mouse doses for the corresponding shifts were recalculated and dose volume histograms (DVHs) were generated. The investigation was performed for 1 cm and 0.5 cm diameter tumours. The DVH resulting from introducing the maximum setup errors for 1 cm diameter tumours showed up to 35% reduced dose to a significant fraction of the tumour volume. The setup errors demonstrated an insignificant effect on doses for 0.5 cm diameter tumour irradiations. Setup errors were observed to have negligible impact on out of field doses to organs at risk. The dosimetric results presented herein verify the robustness of our collimator system for irradiations of xenograft tumours up to 0.5 cm diameter in the presence of the maximum setup errors.

Publisher

Research Square Platform LLC

Reference35 articles.

1. Ford NL, Loudos GK, Karnabatidis DN, Kagadis GC (2018) Defining small animal imaging, therapy, and applications. Handbook of Small Animal Imaging. CRC Press, pp 3–5

2. Milas L, Milas MM, Mason KA (1999), April Combination of taxanes with radiation: preclinical studies. In Seminars in radiation oncology (Vol. 9, No. 2 Suppl 1, pp. 12–26)

3. Combining radiation therapy and cancer immune therapies: From preclinical findings to clinical applications;Bockel S;Cancer/Radiothérapie,2018

4. Combined treatment effects of radiation and immunotherapy: studies in an autochthonous prostate cancer model;Wada S;Int J Radiation Oncology* Biology* Phys,2013

5. Vivo Bioluminescence Tomography Center of Mass-Guided Conformal Irradiation;Deng Z;Int J Radiation Oncology* Biology* Phys,2020