Affiliation:
1. Prirodno-matematički fakultet, Odsek za hemiju, Niš
2. Hemijski fakultet, Beograd
Abstract
A new reaction was suggested and a new kinetic method was elaborated for determination of Bi(III) in solution, based on its catalytic effect on the oxidation of phenyl-fluorone (PF) by hydrogen peroxide in ammonia buffer. By application of spectrophotometric technique, a limit of quantification (LQ) of 128 ng cm-3 was reached, and the limit of detection (LD) of 37 ng cm-3 was obtained, where LQ was defined as the ratio signal: noise = 10:1 and LD was defined as signal 3:1 against the blank. The RSD value was found to be in the range 2.8-4.8 % for the investigated concentration range of Bi(III). The influence of some ions upon the reaction rate was tested. The method was confirmed by determining Bi(III) in a stomach ulcer drug ('Bicit HP', Hemofarm A.D.). The obtained results were compared to those obtained by AAS and good agreement of results was obtained.
Publisher
National Library of Serbia
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献