Author:
Price R. B. T.,Fahey J.,Felix C. M.
Abstract
Clinical Relevance
The output of a curing light measured at a distance of 0 mm is a poor indicator of how much light energy will be delivered to a restoration in the mouth. When used for manufacturers' suggested curing times and at clinically relevant distances, some curing lights deliver much less energy than is recommended (<10J/cm2) and produce softer composites.
Cited by
27 articles.