Affiliation:
1. Department of Electrical Engineering National Chin‐Yi University of Technology Taichung City Taiwan
2. Department of Medicine of Chi Mei Medical Center Chien‐Ming Li is with the Division of Infectious Diseases Tainan City Taiwan
Abstract
AbstractVital signs such as heart rate (HR) and respiration rate (RR) are essential physiological parameters that are routinely used to monitor human health and bodily functions. They can be continuously monitored through contact or contactless measurements performed in the home or a hospital. In this study, a contactless Doppler radar W‐band sensing system was used for short‐range, contactless vital sign estimation. Frequency‐modulated continuous wave (FMCW) measurements were performed to reduce the influence of a patient's micromotion. Sensing software was developed that can process the received chirps to filter and extract heartbeat and breathing rhythm signals. The proposed contactless sensing system eliminates the need for the contact electrodes, electric patches, photoelectric sensors, and conductive wires used in typical physiological sensing methods. The system operates at 76–81 GHz in FMCW mode and can detect objects on the basis of changes in frequency and phase. The obtained signals are used to precisely monitor a patient's HR and RR with minimal noise interference. In a laboratory setting, the heartbeats and breathing rhythm signals of healthy young participants were measured, and their HR and RR were estimated through frequency‐ and time‐domain analyses. The experimental results confirmed the feasibility of the proposed W‐band mm‐wave radar for contactless and short‐range continuous detection of human vital signs.
Publisher
Institution of Engineering and Technology (IET)