2. Department of Biomedical Engineering Kyungpook National University Hospital Daegu South Korea
Abstract
AbstractThis study describes the design of a transducer for bone conduction implants that combines a piezoelectric element and a displacement amplifier. To develop a displacement amplifier with the maximum possible amplification ratio, theoretical analysis was performed to derive the parameters that affect the amplification ratio. Parametric sweep analysis was conducted to calculate the amplification ratio according to changes in the various parameters. The resulting optimal displacement amplifier afforded the maximum amplification ratio when the beam thickness, depth, inclination angle, and length were 0.15 mm, 1.2 mm, 6.5°, and 4 mm, respectively. The displacement amplifier with these optimal parameter values amplified the piezoelectric element displacement by approximately 7.7‐fold. Frequency characteristic analysis by weight was performed to derive a mechanical resonance location with frequency characteristics that appropriately compensated for hearing loss. As a result of this analysis, when a weight of 0.7 g was applied to the displacement amplifier, the mechanical resonance was that required for bone conduction implant transducers.