Intraoperative placement of total hip arthroplasty components with robotic-arm assisted technology correlates with postoperative implant position

Author:

Nodzo S. R.1,Chang C-C.2,Carroll K. M.3,Barlow B. T.4,Banks S. A.5,Padgett D. E.3,Mayman D. J.3,Jerabek S. A.3

Affiliation:

1. Department of Orthopedics, Mike O’Callaghan Medical Center, Las Vegas, Nevada, USA.

2. Department of Mechanical & Aerospace Engineering, University of Florida, Gainesville, Florida, USA.

3. Department of Orthopedics, Hospital for Special Surgery, New York, New York, USA.

4. San Diego Naval Medical Center, San Diego, California, USA.

5. Department of Mechanical and Aerospace Engineering, University of Florida, Gainesville, Florida, USA.

Abstract

Aims The aim of this study was to evaluate the accuracy of implant placement when using robotic assistance during total hip arthroplasty (THA). Patients and Methods A total of 20 patients underwent a planned THA using preoperative CT scans and robotic-assisted software. There were nine men and 11 women (n = 20 hips) with a mean age of 60.8 years (sd 6.0). Pelvic and femoral bone models were constructed by segmenting both preoperative and postoperative CT scan images. The preoperative anatomical landmarks using the robotic-assisted system were matched to the postoperative 3D reconstructions of the pelvis. Acetabular and femoral component positions as measured intraoperatively and postoperatively were evaluated and compared. Results The system reported accurate values for reconstruction of the hip when compared to those measured postoperatively using CT. The mean deviation from the executed overall hip length and offset were 1.6 mm (sd 2.9) and 0.5 mm (sd 3.0), respectively. Mean combined anteversion was similar and correlated between intraoperative measurements and postoperative CT measurements (32.5°, sd 5.9° versus 32.2°, sd 6.4°; respectively; R2 = 0.65; p < 0.001). There was a significant correlation between mean intraoperative (40.4°, sd 2.1°) acetabular component inclination and mean measured postoperative inclination (40.12°, sd 3.0°, R2 = 0.62; p < 0.001). There was a significant correlation between mean intraoperative version (23.2°, sd 2.3°), and postoperatively measured version (23.0°, sd 2.4°; R2 = 0.76; p < 0.001). Preoperative and postoperative femoral component anteversion were significantly correlated with one another (R2 = 0.64; p < 0.001). Three patients had CT scan measurements that differed substantially from the intraoperative robotic measurements when evaluating stem anteversion. Conclusion This is the first study to evaluate the success of hip reconstruction overall using robotic-assisted THA. The overall hip reconstruction obtained in the operating theatre using robotic assistance accurately correlated with the postoperative component position assessed independently using CT based 3D modelling. Clinical correlation during surgery should continue to be practiced and compared with observed intraoperative robotic values. Cite this article: Bone Joint J 2018;100-B:1303–9.

Publisher

British Editorial Society of Bone & Joint Surgery

Subject

Orthopedics and Sports Medicine,Surgery

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3