FRACTURE OF THE METAL TIBIAL TRAY AFTER KINEMATIC TOTAL KNEE REPLACEMENT

Author:

Abernethy P. J.1,Robinson C. M.1,Fowler R. M.1

Affiliation:

1. Princess Margaret Rose Orthopaedic Hospital, 41–43 Frogston Road West, Edinburgh EH10 7ED, UK.

Abstract

We reviewed 1567 elective knee replacements performed between 1980 and 1990, using either the Total Condylar prosthesis with an all-plastic tibial component, or the Kinematic prosthesis which has a metal tibial tray. The ten-year probability of survival was 92.1% for the Total Condylar design and 87.9% for the Kinematic. The difference was mainly due to 16 revisions required in the Kinematic series for fracture of the metal base-plate. This was the most common cause of aseptic failure in this group. These fractures were strongly associated with a preoperative varus deformity (hazard ratio (HR) 8.8) and there was a slightly increased risk in males (HR 1.9) and in osteoarthritic knees (HR 1.8). In the nine fractures which occurred within four years of primary implantation (group 1), failure to correct adequately a preoperative varus deformity and the use of a bone graft to correct such a deformity were both strongly associated with fracture (HR 13.9 and 15.8, respectively). In eight fractures which occurred more than five years after primary replacement (group 2) we could detect no significant risk factors. Early complications occurred in two patients after the 16 revision procedures for tray fracture. One had a deep infection and the other refracture of the tray.

Publisher

British Editorial Society of Bone & Joint Surgery

Subject

Orthopedics and Sports Medicine,Surgery

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3