Simultaneous in vitro measurement of patellofemoral kinematics and forces following Oxford medial unicompartmental knee replacement

Author:

Price A. J.1,Oppold P. T.2,Murray D. W.1,Zavatsky A. B.2

Affiliation:

1. Nuffield Department of Orthopaedic Surgery, Nuffield Orthopaedic Centre NHS Trust, Windmill Road, Oxford OX3 7LD, UK.

2. Department of Engineering Science, University of Oxford, Parks Road, Oxford OX1 3PJ, UK.

Abstract

The Oxford medial unicompartmental knee replacement was designed to reproduce normal mobility and forces in the knee, but its detailed effect on the patellofemoral joint has not been studied previously. We have examined the effect on patellofemoral mechanics of the knee by simultaneously measuring patellofemoral kinematics and forces in 11 cadaver knee specimens in a supine leg-extension rig. Comparison was made between the intact normal knee and sequential unicompartmental and total knee replacement. Following medial mobile-bearing unicompartmental replacement in 11 knees, patellofemoral kinematics and forces did not change significantly from those in the intact knee across any measured parameter. In contrast, following posterior cruciate ligament retaining total knee replacement in eight knees, there were significant changes in patellofemoral movement and forces. The Oxford device appears to produce near-normal patellofemoral mechanics, which may partly explain the low incidence of complications with the extensor mechanism associated with clinical use.

Publisher

British Editorial Society of Bone & Joint Surgery

Subject

Orthopedics and Sports Medicine,Surgery

Reference34 articles.

1. Walker PS. Requirements for successful total knee replacements: design considerations. Orthop Clin North Am 1989;20:15–29.

2. The mechanics of the knee and prosthesis design

3. Banks SA, Harman MK, Hodge WA. Mechanism of anterior impingement damage in total knee arthroplasty. J Bone Joint Surg [Am] 2002;84-A(Suppl 2):37–42.

4. Banks SA, Markovich GD, Hodge WA. The mechanics of knee replacements during gait: in vivo fluoroscopic analysis of two designs. Am J Knee Surg 1997;10:261–7.

5. Patellofemoral forces after total knee arthroplasty: effect of extensor moment arm

Cited by 31 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献