Affiliation:
1. Centre for Hip Surgery, Wrightington Hospital
Abstract
It has been shown by mechanical analysis that by using cement to bond the stem of a femoral head prosthesis to bone two advantages are obtained when the conditions are compared with conventional methods. 1. "Fretting" between the implant and the living bone is eliminated. This source of persistent relative movement is probably the most important starting point for the progressive loosening of weight-bearing implants. 2. When cement is used the bond with the bone is exposed to stresses which are of an order three hundred times less than the shear strength of bone. The conventional prostheses expose the bond to compressive stresses which are near to the failure limits of the compressive strength of bone, especially in elderly patients with atrophic cortical bone in the femoral neck.
Publisher
British Editorial Society of Bone & Joint Surgery
Subject
Orthopedics and Sports Medicine,Surgery
Cited by
79 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献