The reliability and accuracy of measuring anteversion of the acetabular component on plain anteroposterior and lateral radiographs after total hip arthroplasty

Author:

Shin W. C.1,Lee S. M.1,Lee K. W.1,Cho H. J.1,Lee J. S.1,Suh K. T.1

Affiliation:

1. Pusan National University School of Medicine, 20 Geumo-ro, Mulgeum-eup, Yangsan, Gyeongsangnam-do, 626-770, Korea.

Abstract

There is no single standardised method of measuring the orientation of the acetabular component on plain radiographs after total hip arthroplasty. We assessed the reliability and accuracy of three methods of assessing anteversion of the acetabular component for 551 THAs using the PolyWare software and the methods of Liaw et al, and of Woo and Morrey. All measurements of the three methods had excellent intra- and inter-observer reliability. The values of the PolyWare software, which determines version of the acetabular component by edge detection were regarded as the reference standard. Although the PolyWare software and the method of Liaw et al were similarly precise, the method of Woo and Morrey was significantly less accurate (p < 0.001). The method of Liaw et al seemed to be more accurate than that of Woo and Morrey when compared with the measurements using the PolyWare software. If the qualified lateral radiograph was selected, anteversion measured using the method of Woo and Morrey was considered to be relatively reliable. Cite this article: Bone Joint J 2015; 97-B:611–16.

Publisher

British Editorial Society of Bone & Joint Surgery

Subject

Orthopedics and Sports Medicine,Surgery

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3