Finite-element analysis of failure of the Capital Hip designs

Author:

Janssen D.1,Aquarius R.1,Stolk J.1,Verdonschot N.1

Affiliation:

1. Orthopaedic Research Laboratory, Radboud University, Nijmegen Medical Centre, P. O. Box 9101, 6500 HB Nijmegen, The Netherlands.

Abstract

The Capital Hip implant was a Charnley-based system which included a flanged and a roundback stem, both of which were available in stainless steel and titanium. The system was withdrawn from the market because of its inferior performance. However, all four of the designs did not produce poor rates of survival. Using a simulated-based, finite-element analysis, we have analysed the Capital Hip system. Our aim was to investigate whether our simulation was able to detect differences which could account for the varying survival between the Capital Hip designs, thereby further validating the simulation. We created finite-element models of reconstructions with the flanged and roundback Capital Hips. A loading history was applied representing normal walking and stair-climbing, while we monitored the formation of fatigue cracks in the cement. Corresponding to the clinical findings, our simulation was able to detect the negative effects of the titanium material and the flanged design in the Capital Hip system. Although improvements could be made by including the effect of the roughness of the surface of the stem, our study increased the value of the model as a predictive tool for determining failure of an implant.

Publisher

British Editorial Society of Bone & Joint Surgery

Subject

Orthopedics and Sports Medicine,Surgery

Reference22 articles.

1. UK Medical Devices Agency Hazard Notice 9801. 1998.

2. EARLY FEMORAL LOOSENING IN ONE DESIGN OF CEMENTED HIP REPLACEMENT

3. Ramamohan N, Grigoris P, Schmolz W, Chapell AM, Hamblen DL. Early failure of stainless steel 3M Capital femoral stem. J Bone Joint Surg [Br] 2000;82-B (Suppl 1):71.

4. 3M Capital hip arthroplasty

5. The Royal College of Surgeons of England. An investigation of the performance of the 3M Capital Hip System. London: The Royal College of Surgeons of England, 2001.