Abstract
A method is described of measuring radiological subsidence of a femoral prosthesis in relation to the femur after total hip arthroplasty. The method depends on measuring the distance from the tip of the femoral prosthesis to a fixed point in the bone. Subsidence after the use of a conventional design of femoral stem is compared with that after the use of a stem with a dorsal flange (Cobra). A significant reduction in the incidence and amount of subsidence was found when using the dorsal flange. There was also a notable absence of transverse fractures involving the cement near the tip of the stem, which occurred in 26 per cent of the cases using a conventional prosthesis.
Publisher
British Editorial Society of Bone & Joint Surgery
Subject
Orthopedics and Sports Medicine,Surgery
Cited by
120 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献