We studied the quantity and rate of formation of new bone during lengthening of 17 limb segments in 10 patients using dual-energy X-ray absorptiometry (DEXA), ultrasonography and radiography. Whereas new bone was detected by both DEXA and ultrasonography within 1 to 2 weeks of distraction, it was not visible on the radiographs until 4 to 8 weeks. Limb alignment and gap measurement were accurately assessed by DEXA without the need for standard radiographs or scanograms. With ultrasound the distraction gap appeared as an echolucent window which narrowed progressively producing a hyper-reflecting line after which further consolidation could not be assessed. As measured by DEXA the density of the new bone at this stage was approximately 45% of control values and did not represent normal cortication. Whereas ultrasound could be used to identify defects in mineralisation and to determine when to dynamise the fixator system, DEXA could measure the quantity and rate of formation of bone throughout lengthening.