Abstract

The Articular Surface Replacement (ASR) hip resurfacing arthroplasty has a failure rate of 12.0% at five years, compared with 4.3% for the Birmingham Hip Resurfacing (BHR). We analysed 66 ASR and 64 BHR explanted metal-on-metal hip replacements with the aim of understanding their mechanisms of failure. We measured the linear wear rates of the acetabular and femoral components and analysed the clinical cause of failure, pre-revision blood metal ion levels and orientation of the acetabular component. There was no significant difference in metal ion levels (chromium, p = 0.82; cobalt, p = 0.40) or head wear rate (p = 0.14) between the two groups. The ASR had a significantly increased rate of wear of the acetabular component (p = 0.03) and a significantly increased occurrence of edge loading (p < 0.005), which can be attributed to differences in design between the ASR and BHR. The effects of differences in design on the in vivo wear rates are discussed: these may provide an explanation as to why the ASR is more sensitive to suboptimal positioning than the BHR.

Publisher

British Editorial Society of Bone & Joint Surgery

Subject

Orthopedics and Sports Medicine,Surgery

Reference32 articles.

1. Amstutz HC, Le Duff MJ. Background of metal-on-metal resurfacing. Proc Inst Mech Eng H2006; 220: 85–94.

2. No authors listed. Medicines and Healthcare products Regulatory Agency. MHRA medical device alert: recall of DePuy ASR hip replacement implants, 2010 (MDA/2010/069). http://www.mhra.gov.uk/Publications/Safetywarnings/MedicalDevice-Alerts/CON093789 (date last accessed 31 May 2011).

3. No authors listed. National Joint Registry for England and Wales: 6th annual report, 2009. http://www.njrcentre.org.uk/NjrCentre/Portals/0/Sixth%20annual%20NJR%20report.pdf (date last accessed 31 May 2011).

4. Issac GH, Siebel T, Schmalzried TP, et al. Development rationale for an articular surface replacement: a science-based evolution. Proc Inst Mech Eng H2006; 220: 253–268.

5. Birmingham hip resurfacing

Cited by 64 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献