Affiliation:
1. R. K. Hospital, Groningen, The Netherlands.
Abstract
Between February 1975 and August 1976, 195 total hip replacements using Stanmore components were performed; of these, 146 were in 135 patients who had not had previous hip surgery. At review 52 had died, but none of the others was lost to follow-up. Of the 52, two had had a second operation, one for infection and one for recurrent dislocation. In the remaining 83 patients (92 hips) five revisions were necessary: four for aseptic loosening, and one for stem fracture. The remaining 78 patients had little or no pain and little restriction of activity. After a follow-up period of at least nine years, the survival rate of the prosthesis was 95%. There had been migration of the femoral component in five cases and migration of the acetabular cup in one case, but no wear of the acetabular component could be demonstrated.
Publisher
British Editorial Society of Bone & Joint Surgery
Subject
Orthopedics and Sports Medicine,Surgery
Cited by
14 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献