Affiliation:
1. Department of Orthopaedic Surgery, Hospital for Joint Diseases, New York, NY 10003.
Abstract
We examined the effect of the Gamma nail on strain distribution in the proximal femur, using ten cadaver femora instrumented with six unidirectional strain gauges along the medial and lateral cortices. The femora were loaded to 1800 N and strains were determined with or without distal interlocking screws before and after experimentally created two-part and four-part fractures. Motion of the sliding screw and the nail was also determined. Strain patterns and screw motion were compared with previously obtained values for a sliding hip screw device (SHS). The Gamma nail was shown to transmit decreasing load to the calcar with decreasing fracture stability, such that virtually no strain on the bone was seen in four-part fractures with the posteromedial fragment removed; increasing compression was noted, however, at the proximal lateral cortex. Conversely, the SHS showed increased calcar compression with decreasing fracture stability. The insertion of distal interlocking screws did not change the pattern of proximal femoral strain. The Gamma nail imparts non-physiological strains to the proximal femur, probably because of its inherent stiffness. These strains may alter bone remodelling and interfere with healing. Distal interlocking screws may not be necessary for stable intertrochanteric fractures.
Publisher
British Editorial Society of Bone & Joint Surgery
Subject
Orthopedics and Sports Medicine,Surgery
Cited by
177 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献