Ten-year results of a double-heat-treated metal-on-metal hip resurfacing

Author:

Daniel J.1,Ziaee H.1,Kamali A.2,Pradhan C.1,Band T.2,McMinn D. J. W.1

Affiliation:

1. The McMinn Centre, 25, Highfield Road, Birmingham, B15 3DP, UK.

2. Implant Development Centre Smith and Nephew Orthopaedics, Aurora House, Spa Park, Harrison Way, Leamington Spa CV31 2NB, UK.

Abstract

Second-generation metal-on-metal bearings were introduced as a response to the considerable incidence of wear-induced failures associated with conventional replacements, especially in young patients. We present the results at ten years of a consecutive series of patients treated using a metal-on-metal hip resurfacing. A distinct feature of the bearings used in our series was that they had been subjected to double-heat treatments during the post-casting phase of their manufacture. In the past these bearings had not been subjected to thermal treatments, making this a unique metal-on-metal bearing which had not been used before in clinical practice. We report the outcome of 184 consecutive hips (160 patients) treated using a hybrid-fixed metal-on-metal hip resurfacing during 1996. Patients were invited for a clinicoradiological follow-up at a minimum of ten years. The Oxford hip score and anteroposterior and lateral radiographs were obtained. The mean age at operation was 54 years (21 to 75). A series of 107 consecutive hips (99 patients) who received the same prosthesis, but subjected to a single thermal treatment after being cast, between March 1994 and December 1995, were used as a control group for comparison. In the 1994 to 1995 group seven patients (seven hips) died from unrelated causes and there were four revisions (4%) for osteolysis and aseptic loosening. In the 1996 group nine patients died at a mean of 6.9 years after operation because of unrelated causes. There were 30 revisions (16%) at a mean of 7.3 years (1.2 to 10.9), one for infection at 1.2 years and 29 for osteolysis and aseptic loosening. Furthermore, in the latter group there were radiological signs of failure in 27 (24%) of the 111 surviving hips. The magnitude of the problem of osteolysis and aseptic loosening in the 1996 cohort did not become obvious until five years after the operation. Our results indicate that double-heat treatments of metal-on-metal bearings can lead to an increased incidence of wear-induced osteolysis.

Publisher

British Editorial Society of Bone & Joint Surgery

Subject

Orthopedics and Sports Medicine,Surgery

Cited by 65 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Hip resurfacing versus total hip arthroplasty: a systematic review and meta-analysis of randomized clinical trials;International Orthopaedics;2024-08-07

2. Birmingham Hip Resurfacing at 20 years;The Bone & Joint Journal;2023-09-01

3. Hip resurfacing LESSONS FROM THE PAST AND DIRECTIONS FOR THE FUTURE;BONE JOINT J;2023

4. Hip resurfacing;The Bone & Joint Journal;2023-05-01

5. Hip and Knee Replacement;Current Orthopaedic Practice;2021-10-01