Abstract
We have reviewed 85 knees in 71 patients after total-condylar posteriorly stabilised (Insall-Burstein) knee replacement with an average follow-up of five years. Excellent or good results were obtained in 90% with an average maximum flexion of 98 degrees. The four poor results (5%) included two with deep infection, one with patellar dislocation and one with loosening. Four other knees (5%) showed signs of probable tibial loosening, but the patients were asymptomatic, the clinical results had not deteriorated with time and lucent lines had not progressed. Varus alignment of the knee and a varus tilt of more than 2 degrees of the tibial component correlated with the incidence of lucent lines around the tibial implant. No patellar stress fractures were seen but impingement symptoms were present in 20%, although they were troublesome in less than half of them. The virtue of the prosthesis lies in its versatility for use in the severely deformed joint.
Publisher
British Editorial Society of Bone & Joint Surgery
Subject
Orthopedics and Sports Medicine,Surgery
Cited by
117 articles.
订阅此论文施引文献