Affiliation:
1. Department of Orthopaedic Surgery, Leicester Royal Infirmary, Leicester, UK.
Abstract
We treated 39 knees with chronic deficiency of the anterior cruciate ligament by reconstruction using the ABC carbon and polyester prosthetic ligament; 31 (79.5%) were reviewed at an average follow-up of 34 months. There had been four complete failures requiring revision. The remaining 27 were studied in detail. On the Lysholm rating, only 11 knees (41%) had good results with a score of over 76. The mean anterior drawer movement was reduced from 7.6 mm before operation to 5.8 mm at review. The mean difference from the opposite uninjured knee was 3.9 mm before operation, 1 mm (in 21 patients) at mean follow-up of 7.4 months and 2.5 mm (in 27 patients) at 34 months, indicating progressive loss of effect. In our opinion the results are unsatisfactory: we do not recommend the use of this prosthetic ligament.
Publisher
British Editorial Society of Bone & Joint Surgery
Subject
Orthopedics and Sports Medicine,Surgery
Cited by
42 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献