Affiliation:
1. Department of Orthopaedic Surgery, Aarhus Kommunehospital, Denmark.
Abstract
We allocated randomly 27 patients undergoing 28 primary uncemented total hip replacements (THR) to receive prosthetic components of similar design with either plasma-sprayed titanium alloy (Ti) coating (n = 13) or hydroxyapatite (HA) coating (n = 15). After some exclusions, 15 of the patients (15 THR; 7 with HA- and 8 with Ti-coating) were followed by roentgen stereophotogrammetric analysis at 3, 6 and 12 months to measure migration of the femoral component. Twenty-six of the patients (26 THR) were followed clinically and by conventional radiography. All the femoral components had migrated at 3 months. From 3 to 12 months, the migration of Ti-coated components continued whereas the HA-coated components had stabilised. At 12 months there was significantly less migration of the HA-coated components (p < 0.05). The maximum subsidence was 0.2 mm in both groups. The Harris hip score was equal in the two groups preoperatively but at follow-up it was better in the HA-coated group (p < 0.05) and visual analogue scale scores showed that they had less pain (p < 0.05).
Publisher
British Editorial Society of Bone & Joint Surgery
Subject
Orthopedics and Sports Medicine,Surgery
Cited by
181 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献