Why large-head metal-on-metal hip replacements are painful

Author:

Cobb J. P.1,Davda K.1,Ahmad A.1,Harris S. J.1,Masjedi M.1,Hart A. J.1

Affiliation:

1. Imperial College, Charing Cross Campus, Fulham Palace Road, London W6 8RF, UK.

Abstract

Large-head metal-on-metal total hip replacement has a failure rate of almost 8% at five years, three times the revision rate of conventional hip replacement. Unexplained pain remains a feature of this type of arthroplasty. All designs of the femoral component of large-head metal-on-metal total hip replacements share a unique characteristic: a subtended angle of 120° defining the proportion of a sphere that the head represents. Using MRI, we measured the contact area of the iliopsoas tendon on the femoral head in sagittal reconstruction of 20 hips of patients with symptomatic femoroacetabular impingement. We also measured the articular extent of the femoral head on 40 normal hips and ten with cam-type deformities. Finally, we performed virtual hip resurfacing on normal and cam-type hips, avoiding overhang of the metal rim inferomedially. The articular surface of the femoral head has a subtended angle of 120° anteriorly and posteriorly, but only 100° medially. Virtual surgery in a normally shaped femoral head showed a 20° skirt of metal protruding medially where iliopsoas articulates. The excessive extent of the large-diameter femoral components may cause iliopsoas impingement independently of the acetabular component. This may be the cause of postoperative pain with these implants.

Publisher

British Editorial Society of Bone & Joint Surgery

Subject

Orthopedics and Sports Medicine,Surgery

Reference8 articles.

1. No authors listed. National Joint Registry for England and Wales: 7th annual report, 2010. http://njrcentre.org.uk (date last accessed 2 April 2011).

2. Jeffers JR, Roques A, Taylor A, Tuke MA. The problem with large diameter metal-on-metal acetabular cup inclination. Bull NYU Hosp Jt Dis 2009;67:189–92.

3. Failure of the Durom Metasul® Acetabular Component

4. The painful metal-on-metal hip resurfacing

5. No authors listed. Medical Device Alert: ARMTM hip replacement implants manufactured by DePuy International Ltd (MDA/2010/069). http://www.mhra.gov.uk/Publications/Safetywarnings/MedicalDeviceAlerts/CON093789 (date last accessed 23 May 2011).

Cited by 38 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3