Affiliation:
1. Sunnybrook Health Science Centre, Toronto, Ontario, Canada.
Abstract
We reviewed 64 patients in whom 66 acetabula had been reconstructed with either the Muller ring (46) or the Burch-Schneider anti-protrusio cage (20) at a mean follow-up of five years. Five hips had been revised a second time for loosening, all after a Muller ring had been used for a medial segmental defect (2), ungrafted cavitary defects (2) or after resorption of a block graft (1). The use of bone grafts with the implants reduced the incidence of failure from 13% to 6% and of circumferential radiolucent lines at the bone-implant interface from 39% to 2%. The Muller ring is indicated for acetabula with isolated peripheral segmental defects or cavitary defects confined to one or two sectors. The Burch-Schneider cage should be used for medial segmental defects, extensive cavitary defects and combined deficiencies. Defects should be reconstituted with bone graft rather than cement.
Publisher
British Editorial Society of Bone & Joint Surgery
Subject
Orthopedics and Sports Medicine,Surgery
Cited by
146 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献