Affiliation:
1. Portsmouth Hospitals NHS Trust, Queen Alexandra Hospital, Southwick Hill Road, Cosham, Portsmouth PO6 3LY, UK.
Abstract
We reviewed the initial post-operative radiographs of the Trident acetabulum and identified a problem with seating of the metal-backed ceramic liner. We identified 117 hips in 113 patients who had undergone primary total hip replacement using the Trident shell with a metal-backed alumina liner. Of these, 19 (16.4%) were noted to have incomplete seating of the liner, as judged by plain anteroposterior and lateral radiographs. One case of complete liner dissociation necessitating early revision was not included in the prevalence figures. One mis-seated liner was revised in the early post-operative period and two that were initially incompletely seated were found on follow-up radiographs to have become correctly seated. There may be technical issues with regard to the implanting of this prosthesis of which surgeons should be aware. However, there is the distinct possibility that the Trident shell deforms upon implantation, thereby preventing complete seating of the liner.
Publisher
British Editorial Society of Bone & Joint Surgery
Subject
Orthopedics and Sports Medicine,Surgery
Reference14 articles.
1. National Joint Registry for England &
Wales. Second Annual Report, Sept 2005. www.njrcentre.org.uk/documents/reports/NJR_2fullreport.pdf (date last accessed 26th October 2006).
2. Ultra-low wear rates for rigid-on-rigid bearings in total hip replacements
3. Cancer Risk After Metal on Metal and Polyethylene on Metal Total Hip Arthroplasty
4. Mahoney OM, Dimon JH 3rd. Unsatisfactory results with a ceramic total hip prosthesis. J Bone Joint Surg [Am] 1990;72-A:663–71.
5. National Joint Replacement Registry. Annual Report of the Australian Orthopaedic Association, 2005. www.dmac.adelaide.au/aoanjrr/documents/aoanjrreport_2005.pdf (date last accessed 26th October 2006).
Cited by
78 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献