Author:
Hooten JP,Engh CA,Engh CA
Abstract
We investigated the radiographic and clinical course of 31 patients in whom a bulk acetabular allograft had been used during the cementless revision of a total hip replacement. Two patients died and two were lost to follow-up within 24 months, but of the remaining 27 acetabular components, 12 (44%) showed radiographic evidence of instability at a mean of 46 months. Five of these have been revised. In the 12 failures, signs of instability had been noted at an average of 29 months (1 to 60). Failures during the first 24 months were usually due to technical errors, later failures to gradual migration of the cup into the graft. The cups with the greatest amount of their surface supported by grafts were most likely to migrate, but this migration was usually asymptomatic. Screw fixation of the cup, used in 24 cases, appeared to control the mechanism of failure. Femoral head allografts and distal femoral allografts had been used, with failure in 6 of 16, and 6 of 11 respectively; distal femoral allografts were used only for large defects. The insidious course of late cup migration and graft failure necessitates close radiographic follow-up of patients treated with bulk allografts.
Publisher
British Editorial Society of Bone & Joint Surgery
Subject
Orthopedics and Sports Medicine,Surgery
Cited by
155 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献