Affiliation:
1. The Departments of Orthopaedic Surgery and of Metallurgy, University of Manchester
Abstract
1. The factors producing electrolytic corrosion of stainless steels are reviewed, and it is shown how several factors operating together can accentuate corrosion in certain positions in the plated fracture. 2. These factors can be minimised by good metallurgy and good engineering on the part of the manufacturers. 3. When and if materials become metallurgically satisfactory, or a truly inert substitute for metal is discovered, it will be possible to assess the true contribution of infection, faulty operative technique and mechanical factors to the failures of internal fixation. Such a critical assessment of these probably more important factors is at present bedevilled by uncertainty as to the purely physical condition of the materials as supplied to the surgeon.
Publisher
British Editorial Society of Bone & Joint Surgery
Subject
Orthopedics and Sports Medicine,Surgery
Cited by
21 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献