Comparison of Bayer Advia Centaur® immunoassay results obtained on samples collected in four different Becton Dickinson Vacutainer® tubes

Author:

Morovat A1,James TS1,Cox SD1,Norris SG2,Rees MC3,Gales MA1,Taylor RP1

Affiliation:

1. Department of Clinical Biochemistry, John Radcliffe Hospital, Oxford OX3 9DU, UK

2. Department of Haematology, John Radcliffe Hospital, Oxford OX3 9DU, UK

3. Nuffield Department of Obstetrics and Gynaecology, John Radcliffe Hospital, Oxford OX3 9DU, UK

Abstract

Background: Medicines and Healthcare products Regulatory Agency's (MHRA's) Medical Device Alert MDA/2004/048 described bias in some endocrine test results obtained on a few immunoassay platforms, particularly the Bayer Advia Centaur® instrument, when using blood specimens collected into Becton Dickinson (BD) VacutainerTM SSTTMII Advance tubes. As users of BD tubes and the Advia Centaur® instrument, we addressed our concerns about the quality of the results that we had previously reported by undertaking an independent study. Method: We compared the results of 15 immunoassays performed on Bayer Advia Centaur® using blood specimens collected into four different BD Vacutainer® tubes (plain, old and newly released BD SSTTMII Advance, and BD PSTTMII). Results: Compared with plain tubes, old SSTTMII Advance tube results showed no bias for testosterone, CA15-3, follicle-stimulating hormone and folate assays, but gave a positive bias for cortisol and a negative bias for vitamin-B12. Compared with plain tubes, BD PSTTMII tubes gave no significant bias for thyroid function tests, prolactin, parathyroid hormone, and CA125, but gave a negative bias for steroid assays, and a positive bias for gonadotrophins. The results obtained using new BD SSTTMII Advance tubes were generally comparable with those on plain tubes. Conclusions: Only for cortisol did our findings support the bias described by MHRA. Based on our results, apart from vitamin-B12 and possibly cortisol, there may have been no significant influence on clinical decisions as a result of using the old BD SSTTMII Advance specimen tubes. New BD SSTTMII Advance tubes and plain tubes give generally comparable results. BD PSTTMII tubes should not be used for steroid hormone measurements on the Bayer Advia Centaur® instrument.

Publisher

SAGE Publications

Subject

Clinical Biochemistry,General Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3