Near-patient testing of blood glucose using the Bayer Rapidlab 860 analyser in a regional neonatal unit

Author:

Peet Andrew C1,Kennedy David M2,Hocking Michael D3,Ewer Andrew K3

Affiliation:

1. Regional Neonatal Unit, Birmingham Women's Hospital, Birmingham B15 6NH, UK

2. Clinical Chemistry Laboratory Birmingham Women's Hospital Birmingham B15 6NH, UK

3. Regional Neonatal Unit Birmingham Women's Hospital Birmingham B15 6NH, UK

Abstract

Background: We evaluated the use of the glucose biosensor on the Bayer Rapidlab 860 (BR 860) blood gas analyser for neonates and compared it with two laboratory-based methods for measuring blood glucose using the Yellow Springs Instruments 2300 STAT PLUS (YSI) analyser and the Vitros 750 XRC (Vitros) analyser. Methods: Arterial, capillary and venous blood samples were taken from 50 neonates, and blood glucose was measured using all three instruments. Blood pH, PO2, haematocrit and plasma albumin data were also recorded. The effect of varying the degree of haemolysis and the haematocrit was examined using cord blood samples prepared in vitro. Results: Day-to-day imprecision of the BR 860 was 4·90% at a mean glucose concentration of 2·55 mmol/L ( n=60), compared with 3·03% at 2·33 mmol/L ( n=34) for the YSI and 2·64% at 3·52 mmol/L for the Vitros ( n=59). Using Deming regression, the BR 860 results correlated well with results produced on both the YSI ( r=0·962, n=47, P<0·001) and Vitros ( r=0·980, n=46, P<0·001) analysers. Bland-Altman difference plots showed that the BR 860 had a mean of the differences of 0·30 mmol/L compared with the YSI and 0·09 mmol/L compared with the Vitros. The Bland-Altman standard deviation of the differences was 0·46 mmol/L when the BR 860 was compared with the YSI and 0·31 mmol/L when the BR 860 was compared with the Vitros. No significant correlation was found between the BR 860 glucose measurements and haematocrit, degree of sample haemolysis, plasma albumin, blood pH or PO2. Conclusions: The BR 860 analyser provides neonatal blood glucose measurements in the clinical setting of a neonatal unit that compare well with results obtained using laboratory-based methods.

Publisher

SAGE Publications

Subject

Clinical Biochemistry,General Medicine

Cited by 15 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3