Evaluation of four portable self-monitoring blood glucose meters

Author:

Lippi Giuseppe,Salvagno Gian Luca1,Guidi Gian Cesare,Negri Mauro2,Rizzotti Paolo

Affiliation:

1. Sezione di Chimica e Microscopia Clinica, Dipartimento di Scienze Morfologico-Biomediche, Université degli Studi di Verona, Ospedale Policlinico G.B. Rossi, Piazzale Scuro, 10, 37134 Verona, Italy

2. Laboratorio di Chimica Clinica ed Ematologia, Ospedale Civile Maggiore - Azienda Ospedaliera, Verona, Italy

Abstract

Self-monitoring of blood glucose (SMBG) is a cornerstone of diabetes care. However, the effectiveness of any glucose-monitoring programme depends on the ability to integrate SMBG into a program of self-care and therapeutic decision-making. Because the accuracy of SMBG is instrument dependent, we analysed 45 heparinized whole blood specimens using four marketed portable glucose meters to evaluate whether their precision and accuracy would be efficient and safe for clinical use. All measurements were standardized and performed by a single expert health-care professional at the same clinical chemistry laboratory. Results were compared with those obtained on the same plasma samples by the hexokinase method on a secondary reference analyser and further analysed according to the error tolerance criteria and the current American Diabetes Association (ADA), Clinical Laboratory Improvement Amendments (CLIA), and National Committee for Clinical Laboratory Standards (NCCLS) guidelines. The within-run imprecision ranged from 2.2% to 3.2%. Passing and Bablok regression analysis yielded slope values from 0.93 to 1.07 and correlation coefficients between 0.994 and 0.998. When compared with the secondary reference analyser, mean variations were between -4.9% and 14.1%, fulfilling in three out of four cases the 5.5% current desirable analytical quality specifications for total error. Nevertheless, when considering the two standard deviations level of this bias, several results exceeded this limit. Although three out of four devices tested achieved or came closer to the NCCLS C30-A2, CLIA and error tolerance targets, none of them met the current analytical ADA thresholds. Despite the acceptable analytical performances, we demonstrated that standardization and harmonization of results in SMBG have not been fully achieved.

Publisher

SAGE Publications

Subject

Clinical Biochemistry,General Medicine

Cited by 14 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Le proteine glicosilate nel diabete;La Rivista Italiana della Medicina di Laboratorio - Italian Journal of Laboratory Medicine;2013-11-21

2. Higher Accuracy of Self-Monitoring of Blood Glucose in Insulin-Treated Patients in Germany: Clinical and Economical Aspects;Journal of Diabetes Science and Technology;2013-07

3. The Accuracy of Point-of-Care Glucose Measurements;Journal of Diabetes Science and Technology;2012-03

4. Validación de un glucómetro en una unidad de cuidados intensivos;Endocrinología y Nutrición;2012-01

5. Validation of a glucose meter at an intensive care unit;Endocrinología y Nutrición (English Edition);2012-01

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3