Abstract

There are many causes of interference in immunoassays causing erratic patient results. A method-specific interference due to antiruthenium antibodies in Roche free thyroxine (fT4) and free triiodothyronine (fT3) assays has been described previously. As a result, a new generation fT4 assay has been introduced by Roche. We describe six cases of interference due to antiruthenium antibodies, where in four cases interference in the Roche thyroid-stimulating hormone (TSH) assay was found as well. This raised the question as to whether other assays on this platform would also give incorrect results in patients with antiruthenium antibodies. Interference due to antiruthenium antibodies was suspected because of discrepancies between clinical presentation and/or TSH, fT4 and fT3 results. Samples of these six patients were re-analysed in Roche Diagnostics Laboratory, where it was demonstrated that the found discrepancies were indeed caused by interfering antiruthenium antibodies. Subsequently, these patients were asked to donate some blood once more for further evaluation, and three subjects agreed to participate. Their plasma was used to assay 18 analytes on Modular E and on a ruthenium-independent platform. The results were compared taking into account the known differences between distinct methods. As expected, significant interference was found in TSH. Also, in the new generation fT4 assay, ruthenium-induced interference was still present. However, the other assays, both competitive and immunometric, did not show clear interference. We therefore conclude that although antiruthenium antibodies theoretically can interfere in all assays on the Modular E platform, this kind of interference is found in the thyroid hormone assays, without marked interference in the other assays.

Publisher

SAGE Publications

Subject

Clinical Biochemistry,General Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3