Evaluation of two devices for point-of-care testing of haemoglobin in neonatal pigs

Author:

Kutter Annette P N1,Mauch Jacqueline Y23,Riond Barbara4,Martin-Jurado Olga1,Spielmann Nelly3,Weiss Markus3,Bettschart-Wolfensberger Regula1

Affiliation:

1. Section of Anaesthesiology, Equine Department, Vetsuisse Faculty of the University of Zurich, Winterthurerstr. 260, 8057 Zurich, Switzerland

2. Department of Anaesthesia and Perioperative Medicine, Kantonsspital Aarau, Tellstr. 15, 5000 Aarau, Switzerland

3. Department of Anaesthesia, University Children's Hospital, Steinwiesstr. 75, 8032 Zurich, Switzerland

4. Clinical Laboratory, Vetsuisse Faculty of the University of Zurich, Winterthurerstr. 260, 8057 Zurich, Switzerland

Abstract

In veterinary medicine, point-of-care testing (POCT) techniques have become popular, since they provide immediate results and only small amounts of blood are needed. However, their accuracy is controversial. Pigs are often used for research purposes and accurate measurement of haemoglobin (Hb) is important during invasive procedures. The aim of this study was to evaluate two different Hb POCT devices in neonatal pigs. A prospective study with 57 pigs of 3–6 weeks of age, weighing 4.1–6.2 kg (median 5.1 kg) was performed. Fifty-seven blood samples were analysed for Hb using a conductivity-based and a photometrical POCT device and compared with a photometrical reference method. Statistical analysis was performed with Bland–Altman analysis, Spearman correlation and Passing–Bablok regression analysis. Hb values ranged from 32 to 108 g/L (median 80 g/L) using the reference method. The bias of the photometrical method (HemoCue®) to the reference method was −1 g/L, with limits of agreement (LOA) of −7 to 6 g/L. The conductivity-based method (i-STAT®) had a bias of −15 g/L with LOA from −24 to −6 g/L. There was a significant association between protein values and the bias of i-STAT versus CellDyn ( r2 = 0.27, P < 0.05) but not with the bias of HemoCue versus CellDyn ( r2 = 0.001, P = 0.79). The lower the protein values were, the lower the Hb values were measured by the i-STAT. The conductivity-based measurement of Hb constantly underestimated Hb values, whereas the photometrical method demonstrated a better accuracy and is therefore more reliable for on-site measurement of Hb in pigs.

Publisher

SAGE Publications

Subject

General Veterinary,Animal Science and Zoology

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3