Comparison of visibility for four self-expanding nitinol bare stents in vitro

Author:

Wiesinger Benjamin1,Stütz Alexander1,Schmehl Jörg1,Groezinger Gerd1,Claussen Claus Detlef1,Wiskirchen Jakub2

Affiliation:

1. Division of Diagnostic Radiology, Department of Radiology, University of Tübingen;

2. Department of Radiology, Franziskus Krankenhaus, Bielefeld, Germany

Abstract

Background Sufficient radiopacity of stents is a prerequisite for safe interventions and minimization of the radiation dose for the patient and the interventionist. Modern nitinol stents are considered less radiopaque compared to formerly used stents. Purpose To evaluate the objective detection rate (ODR) and the subjective radiopacity score (SRS) of four self-expanding nitinol stents with their markers on a phantom human pelvis. Material and Methods We evaluated the ODR (as a percentage of correctly identified stents) and the SRS (on a scale from 0 = not visible to 4 = excellent visibility) for four self-expanding nitinol stents (SinusSuperflex, SMART, Luminexx, Zilver) with 8 mm diameter and 40 mm length. Stents were placed on a phantom human pelvis and images of the stents were taken in four different positions (right and left lumbosacral joint and near the right and left limbus acetabuli) using the following modes: spotfilm, pulsed fluoroscopy (4, 7.5, 15, and 30 pulses/min) and at three different digital magnification modes. Dose area products (DAPs) were assessed. Results ODR and SRS, respectively, were significantly increased for the SMART stent compared to all other tested stents ( P < 0.05): SMART 93.53% and 2.43, SinusSuperflex 90.81% and 2.21, Luminexx 90.39% and 2.20, and Zilver 89.28% and 2.21. ODR was significantly reduced in position 3 where the bone overlap was more pronounced for all stents (detection rates 77.14-79.56%). An increase in magnification significantly improved the ODR and SRS for all stents (70.33-99.25% and 1.07-3.28, respectively, P < 0.05). Increased pulsing frequency did not improve the ODR of the various stents but did increase the DAP. Conclusion The SMART stent had the best overall performance. In the presence of bone overlap, all self-expanding nitinol stents had poor results. Increased pulsing frequency did not improve ODR or SRS but did increase the DAP. Use of digital magnification modes had no effect on DAP increasing ODR and SRS.

Publisher

SAGE Publications

Subject

Radiology Nuclear Medicine and imaging,General Medicine,Radiological and Ultrasound Technology

Cited by 6 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3