Affiliation:
1. Washington University School of Medicine, St. Louis, MO
Abstract
Purpose
Changes in auditory thresholds following cochlear implantation are generally assumed to be due to damage to neural elements. Theoretical studies have suggested that placement of a cochlear implant can cause a conductive hearing loss. Identification of a conductive component following cochlear implantation could guide improvements in surgical techniques or device designs. The purpose of this study is to characterize new-onset conductive hearing losses after cochlear implantation.
Method
In a prospective study, air- and bone-conduction audiometric testing were completed on cochlear implant recipients. An air–bone gap equal to or greater than 15 dB HL at 2 frequencies determined the presence of a conductive component.
Results
Of the 32 patients with preoperative bone-conduction hearing, 4 patients had a new-onset conductive component resulting in a mixed hearing loss, with air-conduction thresholds ranging from moderate to profound and an average air–bone gap of 30 dB HL. One had been implanted through the round window, 2 had an extended round window, and 1 had a separate cochleostomy.
Conclusions
Loss of residual hearing following cochlear implantation may be due in part to a conductive component. Identifying the mechanism for this conductive component may help minimize hearing loss. Postoperative hearing evaluation should measure both air- and bone-conduction thresholds.
Publisher
American Speech Language Hearing Association
Cited by
29 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献